1-630-784-1000 Fax: 630-784-1651 Page 1 of 4 P/N 1011090 06/10
1-800-253-2991
A. J. ANTUNES & CO. www.ajantunes.com 180 Kehoe Blvd., Carol Stream, Illinois 60188
Dimensions
IMPORTANT
Please read these instructions carefully to assure
correct installation. Prior to being put into opera-
tion, check all wiring and function of the unit.
HLC Installation Information
Wiring Diagram
Operation
When power is applied, the high limit relay will energize
into the normally closed position. If the high limit set point
is exceeded then the relay will de-energize opening the
circuit. The high limit relay can only be reset when the
temperature has dropped below the set point and the
manual reset button (on the front label) is depressed. The
high limit has a visual alarm LED, which is located in the
upper right corner.
Fault Indication: When the LED is “on”, the unit is
latched and in a safe condition (relay open). It can only
be reset when the fault condition is cleared.
Fault Conditions: Over temperature compared to
set point, open or shorted sensor, or any on-board
component failure.
Connections
Thermistor Input: T1, T2 & T3 (100K ohm @ 25°C)
Power Input: T4 & T5 (24-120 VAC)
High Limit Relay: T6 (N.O), T7 (COM)
Installation
For optimum accuracy, the sensor must be mounted with
a minimum of two feet of cable at the sensor lug end
being exposed to the same temperature as the sensor
lug.
5.12”
(130 mm)
2.56”
(65 mm)
3.47”
(88 mm)
2.97”
(75 mm)
2.47”
(63 mm)
3.02”
(77 mm)
BOTTOM VIEW
FULL ENCLOSURE - OPTIONAL
SIDE VIEW
FULL ENCLOSURE - OPTIONAL
.156”
NOTE: Dimensions are for reference only.
T1
T2
T3
T7
T6
T4
T5
Red T1
Black T2
White T3
Sensor
Input
Voltage
24 VAC
COM
N.O.
Output
Relay
T7 T6 T4 T5
120 VAC
Specifications
Sensor Type: 100K Thermistor (Dual) - Length 15 ft.
(4.6 m) Other lengths available upon request.
Accuracy: +/-3ºF of Setpoint
Power Input: 24 VAC, 120 VAC, 60 Hz.
Control Operation: High Limit - Opens on
Temperature Rise
Output Ratings: High Limit - 8 Amp. Max. S.P.S.T. Relay,
240 VAC Max. (Resistive Load)
Adjustable Temperature Range: High Limit: 180º to 300ºF
or 100ºF to 220ºF
Ambient Operating Temperature: -30ºF to 140ºF
(-35ºF to 60ºC)
Connections: 1/4” Quick Connects
Package: Aluminum “C” Bracket
Full Enclosure with Conduit Knockout (optional)
Shipping Weight: 2 lbs. (.9 kg)
Page view 0
1 2 3 4

Summary of Contents

Page 1 - HLC Installation Information

1-630-784-1000 Fax: 630-784-1651 Page 1 of 4 P/N 1011090 06/101-800-253-2991A. J. ANTUNES & CO. www.ajantunes.com 180 Kehoe Blvd., Carol

Page 2

A. J. ANTUNES & CO. www.ajantunes.com 180 Kehoe Blvd., Carol Stream, Illinois 60188HLC Installation Information1-630-784-1000 Fax: 630-784-

Page 3 - 1-800-253-2991

A. J. ANTUNES & CO. www.ajantunes.com 180 Kehoe Blvd., Carol Stream, Illinois 601881-630-784-1000 Fax: 630-784-1651 Page 3 of 4 P/N 10110

Page 4

A.J. Antunes & Co. Headquarters/Manufacturing180 Kehoe BoulevardCarol Stream, Illinois 60188 USAPhone (630) 784-1000Toll Free (800) 253-2991Fax: (

Comments to this Manuals

No comments